pH/mV/°C meters, bench top, Orion™ Versa Star Pro™

Supplier: Thermo Orion
VSTAR12 VSTAR10
ORIOVSTAR12EA 0 SGD
ORIOVSTAR12 ORIOVSTAR10
pH/mV/°C meters, bench top, Orion™ Versa Star Pro™
Electrochemistry Meters pH and Multi-Parameter Meters
These bench top meters with conductivity module meet most challenging applications for pH, mV, ORP and temperature. The meters offer interchangeable measurement modules that allow multiple users to customise four separate channels to meet their specific requirements.

  • Colour display with adjustable text sizes; numeric keypad; onscreen measurement stability indicator
  • Auto-Read to hold a stable reading; ten password-protected methods per channel
  • Audible alarms automatically notify users when a set offset value, high/low limit value, and calibration time
More Product Information
Order Now

SPECIFICATIONS

Model Orion™ Versa Star™ Pro™, with pH module Orion™ Versa Star™ Pro™, with pH module
Environmentally Preferable
pH range −2,000...+20,000 −2,000...+20,000
pH resolution 0,1/0,01/0,001 0,1/0,01/0,001
pH accuracy ±0,002 ±0,002
pH calibration Up to 6 points Up to 6 points
mV range ±2000,0 ±2000,0
mV resolution 0,1 0,1
mV accuracy ±0,2 or 0,05% of reading (whichever is greater) ±0,2 or 0,05% of reading (whichever is greater)
ISE range -
ISE accuracy -
ISE calibration -
ISE units -
Conductivity range -
Conductivity resolution -
Conductivity accuracy -
Resistivity range -
Salinity range -
TDS range -
TDS conversion factor -
DO concentration -
DO concentration resolution -
DO saturation -
DO accuracy -
DO calibration -
Temperature range −5,0...+105,0 °C −5,0...+105,0 °C
Temperature resolution 0,1 °C 0,1 °C
Temperature accuracy ±0,1 °C ±0,1 °C
Temperature calibration 1 point offset 1 point offset
Temperature compensation Manual or automatic with ATC temperature probe Manual or automatic with ATC temperature probe
Display Colour graphic LCD with backlight Colour graphic LCD with backlight
Compliance IP 54 IP 54
Data logging - log function Automatic data logging with Auto-Read, Timed and Single Shot measure modes; manual data logging with Continuous measure mode Automatic data logging with Auto-Read, Timed and Single Shot measure modes; manual data logging with Continuous measure mode
Data logging capacity 2000 points 2000 points
Connections BNC (pH or ORP electrode), pin-tip (reference electrode), 8 pin MiniDIN (ATC temperature probe), two stirrer inputs (stirrer probe) BNC (pH or ORP electrode), pin-tip (reference electrode), 8 pin MiniDIN (ATC temperature probe), two stirrer inputs (stirrer probe)
Outputs RS232, USB (2) RS232, USB (2)
W×D×H 280×240×100 mm 280×240×100 mm
Weight 1,2 kg (modules add 0,2 kg each) 1,2 kg (modules add 0,2 kg each)