pH/mV/°C/ISE meter, bench, Orion™ Dual Star™

Supplier: Thermo Orion
2115000 2115001 2115101 2115102
ORIO2115000EA 0 SGD
ORIO2115000 ORIO2115001 ORIO2115101 ORIO2115102
pH/mV/°C/ISE meter, bench, Orion™ Dual Star™
Electrochemistry Meters pH and Multi-Parameter Meters
These bench top meters with two measuring channels are designed for accurate and reliable pH, ion concentration, mV, ORP and temperature measurements. Ideal for high throughput pH and ion analysis in the lab.

  • Numeric keypad and menu-specific function keys; split screen display for viewing two measurements at the same time
  • Perform up to a 6 point pH calibration with automatic recognition for USA/NIST and EURO/DIN buffers with option to manually input custom buffer values
  • Perform up to a 6 point ion concentration (ISE) calibration with non-linear auto-blank
More Product Information
Order Now

SPECIFICATIONS

Model Dual Star™
Environmentally Preferable
pH range −2,000...+19,999
pH resolution 0,1, 0,01, 0,001
pH accuracy ±0,002
pH calibration 1 - 6 points with automatic buffer recognition (NIST, DIN, custom)
mV range ±1999,9
mV resolution 0,1
mV accuracy ±0,2 of measured value or 0,05% of reading (whichever is greater)
ISE range 0 - 19999
ISE resolution 1, 2 or 3 significant digits
ISE accuracy ±0,2 of measured value or 0,05%
ISE calibration 2 - 6 points (linear, point to point, non-linear selectable auto-blank)
ISE units ppm, ppb, M, % or unitless
Temperature range –5...+105 °C
Temperature resolution 0,1 °C
Temperature accuracy ±0,1 °C
Temperature calibration 1 point temperature offset calibration
Temperature compensation Manual or automatic with ATC temperature probe
Display Graphic LCD with backlight
Power supply Universal AC adapter (included)
Compliance IP 54
Operating temperature range 5…45 °C
Data logging - log function Automatic data logging with auto-read, on ready, at timed intervals, value change in measurement and timed reading measure modes; manual data logging with continuous measure mode
Data logging capacity 1000 points with time/date stamp
Memory 20 methods (10 per channel with calibration data)
Languages English, Spanish, French, Italian, and German
IP rating IP 54
Dimensions 24L×17W×9H cm (9.5×6.7×3.5")
Connections Two BNC (pH, ORP or ion selective electrode), two pin-tip (reference electrode), two 8 pin MiniDIN (ATC temperature probe), stirrer input (stirrer probe)
Outputs USB, RS232
Input Two BNC (pH, ORP or ion selective electrode), two pin-tip (reference electrode), two 8 pin MiniDIN (ATC temperature probe), stirrer input (stirrer probe)
Alarm High/low limit alarm
Parameters pH, mV, relative mV (RmV), ORP or ion concentration (ISE) with temperature
Calibration pH with calibration editing option, ion concentration (ISE) with calibration editing option, relativ.
Calibration points
W×D×H 240×170×90 mm
Weight 0,9 kg

Learn more

About VWR

Avantor is a vertically integrated, global supplier of discovery-to-delivery solutions for...

Learn more About VWR